Drug vacuum freezing dryer clean:
1.
Usually use demineralized water or distilled water as a cleaning agent.
Do not use with acid or base detergents or cleaner, because no matter how cleaning demineralized water, may still leave traces of detergent, they increase the risk of corrosion or a powder into the products.
Clear effect increases with cleaner temperature, but if have a protein content of pollutants, the temperature should not be higher than 60 ℃.
2.
Cleaning method can be used the brush manual cleaning, spray gun manual injection, mainly USES the CIP online cleaning first dense mass of automatic cleaning system.
It is made of many nozzle installed inside a drying chamber and water vapour condenser.
Suction cleaner water mercury, pipelines, valves from nozzle wash drying oven and the surface of the water vapor condenser internal parts, to take away contaminants and impurities, water from the drain discharge at the bottom of the container.
Generally speaking, the clean water is not reusable because cleaning cycle is in danger of comprehensive pollution.